Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: ECD 5100: (180 day) CAL VER /5500 DATE: 02-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 5 NUMBER OF LINES: 95 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Connect UUT to AC power and turn unit ON. 1.006 DISP Warm-up time: 15 minutes 1.006 DISP Temperature: 0 -50degC 1.006 DISP Humidity: 30% - 60% 1.007 DISP [32] WARNING 1.007 DISP HIGH VOLTAGE is used or exposed during the performance 1.007 DISP of this calibration. DEATH ON CONTACT may result if 1.007 DISP personnel fail to observe safety precautions. 1.008 DISP This procedure was written to test the 5000-series 1.008 DISP dataloggers with a TV-10 or TV-40 MUX CARD installed. 1.009 HEAD MAX CHANNEL CONFIGURATION 1.010 DISP NOTE When programming data into the 5000-series 1.010 DISP datalogger, be certain that a "beep" is heard when 1.010 DISP pressing a key to insure the CPU has responded to 1.010 DISP the key stroke. 1.011 DISP Press the key to select alternate. 1.011 DISP Press the <9> key to select the MAX channels. 1.011 DISP Press the <0>, then <2> to select two channels. 1.012 HEAD DWELL CONFIGURATION 1.013 DISP Press the key to select alternate. 1.013 DISP Press the <6> key to select the dwell time. 1.013 DISP Press the <0>, then <5> to set dwell to 5sec. 1.014 HEAD RESET M & B PROGRAMMING CONSTANTS 1.015 DISP Press the key to select alternate. 1.015 DISP Press the key to select M coefficient. 1.015 DISP Press the <+>,<1>,<0>,<0>,<0>,<0> keys to set the 1.015 DISP M constant to +1.0000. 1.016 DISP Press the key to select alternate. 1.016 DISP Press the key to select B coefficient. 1.016 DISP Press the <+>,<0>,<0>,<0>,<0>,<0> keys to set the 1.016 DISP B constant to +0.0000. 1.017 HEAD SET CHANNEL 1 TO READ VOLTAGE 1.018 DISP Press the key to select alternate. 1.018 DISP Press the key to select voltage. 1.018 DISP Press the <0>, then <1> to set the channel to 1. 1.019 HEAD SET CHANNEL 2 TO READ THERMOCOUPLE (J-type) 1.020 DISP Press the key to select alternate. 1.020 DISP Press the key to select J-type thermocouple. 1.020 DISP Press the <0>, then <2> to set the channel to 2. 1.021 HEAD SET CHANNEL 2 TADJUST 1.022 DISP Press the key to select alternate. 1.022 DISP Press the key to select Tadjust. 1.022 DISP Press the <+>,<0>,<0>,<0> keys to set the 1.022 DISP Tadjust to +0000.0. 1.023 HEAD {DC VOLTAGE} 1.024 DISP Connect the 5500A to the TV-10(TV-40)MUX as follows: 1.024 DISP [32] 5500A NORMAL HI to CHANNEL 1 HI 1.024 DISP [32] 5500A NORMAL LO to CHANNEL 1 LO 1.025 DISP HOLD the datalogger on channel 1 by pressing 1.025 DISP the ,,<0>,<1> keys. 1.026 5500 2.0000V 0.0004U 2W 2.001 HEAD {TEMPERATURE} 2.002 DISP HOLD the datalogger on channel 2 by pressing 2.002 DISP the ,,<0>,<2> keys. 2.003 DISP Connect the 5500A TC terminals to the TV-10(TV-40)MUX 2.003 DISP TC INPUT using J type thermocouple wire and connectors. 2.004 M550 TCREF 70degF 2.005 5500 -200.0degF 5U _J TC 2W 3.001 5500 200.0degF 5U _J TC 2W 4.001 5500 1200.0degF 5U _J TC 2W 5.001 END